DIAGNOSTICS Using the Diagnostics
10DEC05
DG4825-1
Page
10 of 180
Section 1: Using the Diagnostics
System Service Diagnostic Screen
Important
For systems with Software Version 4.1, you must have a “Session ID” for access to
“Service Functions” and “Diagnostics”. See SERVICE BULLETIN 843.
1 At the main menu, touch:
[Service Functions]
[ENTER]
[Diagnostics]
2 Does the “System Service Diagnostics” screen display?
3 Check that the TX and RX LEDs are blinking on the MASTER CENTRAL PROCESSING
UNIT BOARD A2 and on the NETWORK CARD in the computer.
4 At the main menu, touch [Service Functions].
5 At the main menu, touch:
[Service Functions]
[ENTER]
[EXIT TO DESKTOP]
6 Select My Computer>Control Panel>Network>Protocols.
7 Double-click [TCP/IP Protocol].
8 Select [ADAPTER 1].
Yes No
Advance to the test or procedure. Continue with Step 3.