DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
98 of 180
16013 - Imaging System contains out-of-range noise
16014 - Imaging System did not zero or PMT did not calibrate
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Description: The “peak-to-peak” signal was high when the CR 825/850 SYSTEM was
set to “AUTO-ZERO.
Cause 1: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Cause 2: Noise in the POWER SUPPLY PS1.
Action: Do the checkout for the POWER SUPPLY PS1.
Cause 1: The DOSIMETER malfunctioned.
Action: Check that the DOSIMETER operates correctly.
Cause 2: The exposure on the PLATE is not correct during the calibration.
Action: Do the calibration with another PLATE.
Cause 3: The X-RAY TUBE malfunctioned.
Action: Expose the PLATE using a new X-RAY TUBE.

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals