DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
114 of 180
47952 - Display only - no Log Message
47960 - Display only - no Log Message
47961 - Display only - no Log Message
47962 - Display only - no Log Message
Cause: Malfunction of the BATTERY.
Action: Do the checkout for the UNINTERRUPTED POWER SUPPLY UPS1.
Description: The UNINTERRUPTED POWER SUPPLY UPS1 has provided
information about a malfunction.
Cause: The UNINTERRUPTED POWER SUPPLY malfunctioned.
Action: Do the checkout for the UNINTERRUPTED POWER SUPPLY UPS1.
Description: The utility power was restored.
Cause: Power is restored.
Action: None.
Description: The power of the BATTERY is full in the UNINTERRUPTED POWER
SUPPLY UPS1.
Cause: BATTERY charge is restored.
Action: None.
Description: The UNINTERRUPTED POWER SUPPLY UPS1 is no longer in a
“FAULT” mode.
Cause: The FAULT mode has cleared.
Action: None.