DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
86 of 180
13010 - Erase time exceeds maximum allowed
13011 - Command to Erase did not occur
13012 - MSC BOARD A1: Invalid Command for ERASE LAMPS
13013 - MSC BOARD A1: Software failure during erase
13899 - General Erase Error
POWER SUPPLY
14001 - SLOW SCAN: Malfunction of 24 V Power
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The software did not provide the RTXC TIMER during the ERASE
function.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The 24 V DC CIRCUIT BREAKER of the SLOW SCAN is open.
Cause: Malfunction of the 24 V DC power distribution.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals