DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
89 of 180
DATA PATH
15001 - MCPU BOARD A2: Fatal Error in Software
15002 - MCPU BOARD A2: Fatal Error in Software
15003 - MCPU BOARD A2: Communications Error in Software
15005 - LASER: Reference Error
Description: A PSOS system error occurred that it is not possible to repair. The
“Event Log” includes the PSOS system error number, module, and line
number.
Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Description: A system software error occurred. The “Event Log” will include the PSOS
system error number, module, and line number.
Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Description: A timeout occurred for a communication between the MASTER
CENTRAL PROCESSING UNIT BOARD A2 and the MOTION SYSTEM
CONTROL BOARD A1.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The reference voltage of the LASER is less than 50% of the original
value.