DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
88 of 180
14006 - DUPLEX CAM: Initialization Error due to the PLATE PRESENT
SENSOR S5 reading blocked when the SLOW SCAN AY is in the home
position
14008 - SLOW SCAN: Initialization Error
14009 - DUPLEX CAM: Initialization Error
14010 - Slow Scan Async Reset Error
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
has occurred. This error occurs when initializing, either when the CR 825/850
SYSTEM is energized or when a mechanical subsystem initializes.
Action: Check:
PLATE was removed from the CASSETTE
SLOW SCAN MOTOR operates correctly
obstruction in the path of the EXTRACTION BAR
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action: Do the checkout PLATE PRESENT SENSOR S5.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause: The CR 825/850 SYSTEM sent a message to reset the SLOW SCAN AY.
Action: None.

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals