DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
104 of 180
22003 - MSC BOARD A1: Checksum of Main Application failed
22004 - MSC BOARD A1: RAM Test failed
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
3. If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
Description: A checksum error occurred in the software of the MOTION SYSTEM
CONTROL BOARD A1. The MOTION SYSTEM CONTROL BOARD A1
remains in the BOOTSTRAP PROGRAM waiting for the software to
download.
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
3. If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
Description: A failure occurred during the test of the RAM on the MOTION SYSTEM
CONTROL BOARD A1.
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
3. If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals