DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
105 of 180
22005 - MSC BOARD A1: Error during download of MSC Software
22006 - MSC BOARD A1: Error during download of MSC Software
22007 - MSC BOARD A1: Error during download of MSC Software
22008 - MSC BOARD A1: Error during download of MSC Software
Description: The boot of the MOTION SYSTEM CONTROL received a signal that is
not correct during the downloading of the software to the MOTION
SYSTEM CONTROL BOARD A1.
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
3. If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
Description: The receive buffer of the boot program received excessive data.
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
3. If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The boot program for the MOTION SYSTEM CONTROL BOARD A1
started to send a message on an RS-232 channel that is not correct.

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals