DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
106 of 180
22009 - MSC BOARD A1: Error during download of MSC Software
22010 - MSC BOARD A1: Error during download of MSC Software
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
3. If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
Description: The boot program for the MOTION SYSTEM CONTROL BOARD A1
received a signal that is not correct.
Cause: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
3. If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
Description: The boot program for the MOTION SYSTEM CONTROL BOARD A1
received a command that was not correct.
Cause 1: The software did not download.
Action: 1. Download the software again.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 3: Communications did not occur between:
MOTION SYSTEM CONTROL BOARD A1
MASTER CENTRAL PROCESSING UNIT BOARD A2
Action: Check the RS-232 communications between:
MOTION SYSTEM CONTROL BOARD A1
MASTER CENTRAL PROCESSING UNIT BOARD A2

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals