DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
83 of 180
12914 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12915 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12930 - SLOW SCAN: Unexpected Retry during Current Test
12931 - SLOW SCAN: Unexpected Retry during Velocity Test
Description: The command MOVE TO EXTRACTION BAR REFERENCE SENSOR
did not occur. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The command MOVE TO THE PLATE GUIDE POSITION did not occur.
The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The system was not able to move to the position selected in the
diagnostics. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The system was not able to move to the position selected in the
diagnostics. The command is executed again
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals