DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
80 of 180
12896 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Reset
12901 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12902 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12254 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.
Error “EM54 - Flash Program Error: Sequence.“
12255 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.
Error “EM55”
12260 Error recording the parameters to the SLOW SCAN CONTROLLER BOARD A6.
Error “EM60 - Parameters Save Flash Error.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The PLATE was not scanned. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The SLOW SCAN AY did not move to the ERASE position. The
command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Code Description

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals