DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
90 of 180
15006 - Communications Error between MCPU BD A2 and GALVO BD A4
Cause 1: The CABLE between the LASER DRIVER BOARD A17 and the GALVO BOARD
A4 malfunctioned.
Action: Check the CABLE.
Cause 2: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.
Cause 3: The LASER DRIVER BOARD A17 malfunctioned.
Action: Do the checkout for the LASER DRIVER BOARD A17.
Description: The DMA process for the MASTER CENTRAL PROCESSING UNIT
BOARD A2 did not complete within the expected time.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3.
Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.