DIAGNOSTICS Using the Diagnostics
10DEC05
DG4825-1
Page
16 of 180
Table 4 Tests
BUTTON Description
[Light Test]
Important
For the test procedure, see Light Test.
Checks for excessive light to the PMTs during the
loading of a CASSETTE.
Allows you to:
actuate or deactuate the high voltage of the PMTs
“AUTO-ZERO” the PMTs
Tests the signal for noise.
[Slow Scan Current and
Velocity Tests]
Caution
An increase in current might indicate that a bind is
occurring in the SLOW SCAN AY.
For the test procedure, see Slow Scan Current and
Velocity Tests.
Drives the SLOW SCAN AY to the limits in both
directions.
Displays for the SLOW SCAN AY:
speed variation
current
“Start of Scan”
“End of Scan”