DIAGNOSTICS Using the Diagnostics
10DEC05
DG4825-1
Page
17 of 180
[Laser Power]
Important
For the test procedure, see Laser Power.
Allows you to:
• check the power of the LASER DIODE
• set the power values for a new LASER DIODE
[Galvo Test]
Important
For the test procedure, see Galvo Test.
Allows you to:
• set the “Amplitude” of the GALVO
• check the response of the LASER DIODE
[PMT Test]
Important
For the test procedure, see PMT Test.
Checks the function of the 2 PMTs.
[Galvo Plot]
Important
For the test procedure, see GALVO PLOT.
Allows you to:
• check:
– response of the GALVO
– automatic gain control - AGC
– 2.5 V reference
• set the GALVO without an OSCILLOSCOPE
BUTTON Description