KV3000 Kinematic Viscosity Bath
Operation and Instruction Manual
KV3000-Manual
-19-
17. Once the bath jar and top assembly are
clean and dry, re-assemble the instrument
and fill the bath with new medium as per
section 5.1 of this document.
7.4 Changing Bath Lamp / Bulb
1. Using a flat head screwdriver, remove the
four (4) silver screws located at the four
corners of the back panel.
2. Remove the back panel plate and set aside.
3. The lamp / light bulb is located at the top
right cover of the instrument. See Figure 5.
Figure 5. Lamp Location
4. Turning counterclockwise, remove the old
light bulb.
5. Replace with new light bulb and re-
assemble the back plate.
7.5 Replacement Parts
Solid State Relay, 4-32 VDC, 20A
RTD Temperature Safety Limit,
120V
†
RTD Temperature Safety Limit,
230V
‡
Slow-Blow Fuse, 1A, 5x20 mm
Slow-Blow Fuse, 0.25A, 5x20 mm
Slow-Blow Fuse, 3.15A, 5x20mm
‡
Time Delay Fuse, 20A, 600VAC
Fast-Acting Fuse, 0.5A, 2AG
Temperature Controller 100-
240V, with RS-232 Board
Compact Fluorescent, 120V, 15W
Stainless Steel Shaft Motor,
115V, 60Hz, 1/20HP
†
Stainless Steel Shaft Motor,
230V, 50/60W, 1/15HP
‡
DPST Contactor, 120V, 30A
†
DPST Contactor, 240V, 30A
‡
Note:
†
For 115V model ONLY (K23700, K23706)
‡
For 220V model ONLY (K23790, K23796)