KV3000 Kinematic Viscosity Bath
Operation and Instruction Manual
KV3000-Manual
-23-
9 Troubleshooting
WARNING: Troubleshooting procedures
involve working with high voltages and/or
temperatures which may result in personal injury
or death, and should only be performed by
trained personnel. Please do not hesitate to
contact Koehler for assistance.
9.1 Unit does not power up
Establish that the socket outlet is providing
proper and adequate voltage.
Check if Overpower Protection circuitry
located directly behind the temperature
controller inside the front tray has been
activated.
Check if line switch is in the ON position.
If problem persists, please call the Koehler
technical service department for assistance.
9.2 Unit is on and keeps resetting into
start up routine
For 220V units, make sure that the socket
outlet is greater than 215V.
Check if there is a steady and reliable power
source.
Make sure the connector plug on the rear
panel is firmly plugged in.
9.3 Unit is on but bath does not heat up
Make sure that the actual temperature
reading is not higher than the set-point
temperature.
Determine if the temperature controller is
properly calibrated by comparison to an
ASTM standard thermometer.
Determine if the Overtemperature Protection
(OTP) circuitry as been activated.
9.4 Bath heats up but temperature does
not stabilize
Make sure there are no drafts from open
doors, windows, or environmental control
vents in the vicinity of the bath.
If the set temperature is close to ambient,
then you may need to circulate cold water
through the coils in order to achieve proper
temperature stability.
10 Service
Under normal operating conditions and with
routine maintenance, the KV3000 Kinematic
Viscosity System should not require service. Any
service problem can be quickly resolved by
contacting Koehler’s technical service
department either by letter, phone, fax, or email.
In order to assure the fastest possible service,
please provide us with the following information.
Model Number:
Serial Number:
Date of Shipment:
11 Storage
This laboratory test instrument is equipped
with electrical components. Storage facilities
should be consistent with an indoor laboratory
environment. This testing equipment should not
be subjected to extremes of temperature and/or
moisture.
This equipment was shipped from the factory
in a corrugated cardboard container. If long term
storage is anticipated, re-packing the instrument
in a water-resistant container is recommended
to ensure equipment safety and longevity.
12 Disposal
12.1 General Recycling Information
No refrigerants or pressurized materials.
No charged capacitors or components that
could electrically discharge.
No components such as springs or spring
powered gears that could store mechanical
energy.
No chemical hazards from any components.
No radiation is emitted from any
components.
12.2 Disposal Information
The KV3000 Kinematic Viscosity Bath is Rohs
compliant and is subject to the WEEE directive.
Once the instrument has reached the end of its
useful life, the instrument and or its components
must be recycled or disposed of in accordance
to Country, State, or local laws that may impose
regulatory requirements regarding disposal.
Dispose instrument and/or components in
accordance to all applicable regulations.
The KV3000 is composed of the following major
components and materials: