What to do if Konica Minolta CM-36dGV show abnormal reflection (color) or gloss results?
- BbishopgabrielleAug 5, 2025
If you're seeing unusual readings for reflection (color) or gloss, here are some things to check: * Make sure your sample is making good contact with the target mask surface. * Verify the target mask is correctly installed. * Clean the inside of the integrating sphere using a blower. If that doesn't work, try zero and white calibration. * Ensure the data for the white calibration plate and gloss calibration plate are correctly set. * Perform zero calibration, white calibration, and gloss calibration correctly.