SPECTROPHOTOMETER
CM
-
3600A
INSTRUCTION MANUAL
Before using this instrument,
please read this manual.

Table of Contents

Questions and Answers

  • T
    Theresa DominguezAug 2, 2025
    Why are reflectance measurement results strange Konica Minolta CM-3600A?
    • S
      Shannon WalshAug 2, 2025
      If reflectance measurement results seem strange on your Konica Minolta Measuring Instruments, first, open the viewfinder cover and check the measurement point. Then, ensure you've set the correct white calibration data. Also, make sure you attach the White Calibration Plate and Zero Calibration Plate correctly, and perform white and zero calibrations correctly. Finally, ensure that there is nothing in the transmittance chamber and perform the operation correctly by referring to the software operation manual.
  • R
    Richard McmillanAug 13, 2025
    What to do if transmittance measurement results are strange Konica Minolta Measuring Instruments?
    • O
      orogersAug 13, 2025
      If you are getting strange transmittance measurement results with your Konica Minolta Measuring Instruments, check the following: * Ensure the specimen is positioned correctly. * Attach the White Calibration Plate correctly and perform 100% calibration correctly using the appropriate calibration standard (air or distilled water). * Attach the White Calibration Plate correctly, block all light from reaching the receptor window, and perform zero calibration correctly. * Refer to the software operation manual to perform the operation correctly.
  • R
    riceedgarAug 23, 2025
    How to fix data input/output issues Konica Minolta CM-3600A Measuring Instruments?
    • J
      Jimmy JohnsonAug 23, 2025
      If data input/output between your Konica Minolta Spectrophotometer and a computer cannot be performed, or if commands cannot be input to the Spectrophotometer from a computer, ensure the USB cable is correctly connected between the Spectrophotometer and the computer. Also, set the POWER switch of Spectrophotometer to OFF and then set it back to ON. Finally, refer to the software operation manual to perform the operation correctly.

Summary

Safety Precautions

Using the CM-3600 A

Measurement Procedure

Flow of Preparation and Measurement

Step-by-step guide for preparing and performing measurements with the CM-3600A.

Connecting a Personal Computer

Connecting the AC Adapter

Turning Power ON and OFF

Attaching a Target Mask

Notes on Use of Target Mask

Important notes regarding the handling and maintenance of target masks.

Attaching the Zero Calibration Box

Notes on Use of Zero Calibration Box

Guidelines for proper use and care of the zero calibration box.

Attaching the White Calibration Plate

Attaching the Transmittance Zero Calibration Plate (Optional Accessory)

Performing 100% calibration

Setting up instrument for 100% calibration to air

Procedure for setting up the instrument for 100% air calibration.

Setting up instrument for 100% calibration to water

Procedure for setting up the instrument for 100% water calibration.

Setting a Specimen

Reflectance measurement

Transmittance measurements

Cleaning the CM-3600 A and Accessories

Error Message

TROUBLESHOOTING GUIDE

Illumination and Measurement Areas

Overview of Simultaneous Measurement of SCI;SCE

Fluorescent Measurement

WHEN FLUORESCENT CALIBRATION IS PERFORMED

Four calibration methods for accurate fluorescent reflectance measurement.

CALCULATING FLUORESCENT REFLECTANCE

How fluorescence is calculated using measured reflectances.

WHEN FLUORESCENT CALIBRATION IS NOT PERFORMED

Measuring fluorescent reflectance without specific calibration.

UV CUT LIGHT SOURCES

Details on available UV cut light sources for fluorescent measurements.

Konica Minolta CM-3600A Specifications

General IconGeneral
ModelCM-3600A
Light SourcePulsed Xenon lamp
Spectral Range360 nm to 740 nm
Wavelength Pitch10 nm
InterfaceUSB, RS-232C
TypeSpectrophotometer
Measurement Geometryd/8°
Integrating Sphere Diameter150 mm
Color SpacesHunter Lab, Munsell
Observer2°/10°
IlluminantD65, A, C
PowerAC 100-240V, 50/60 Hz
Measurement AreaMAV: Ø8 mm / SAV: Ø3 mm
RepeatabilityΔE*ab 0.03 (MAV, white tile)
Colorimetric DataCIE XYZ, CIE L*a*b*, CIE L*C*h