Spectrophotometer
CM-2600d/2500d
E
Instruction Manual
Es
Manual de instrucciones

Table of Contents

Other manuals for Konica Minolta CM-2600d

Questions and Answers

  • L
    Lisa MartinezAug 2, 2025
    What to do if Konica Minolta CM-2600d show ER02 LOW BATTERY?
    • M
      michellelopezAug 2, 2025
      If your Konica Minolta Measuring Instruments display the error 'ER02 LOW BATTERY', it means the battery voltage has dropped. Replace the batteries with new ones to resolve this issue.
  • L
    Lisa DavenportAug 8, 2025
    What to do if Konica Minolta CM-2600d say ER01 MEMORY FULL?
    • D
      Donna BeckerAug 9, 2025
      If your Konica Minolta Measuring Instruments show 'ER01 MEMORY FULL', delete some data to free up memory space. This instrument can store up to 1700 pieces of data in total (700 in the 'defined in COND.' mode), including sample data for the COND and TASK modes, as well as color difference target data.
  • Z
    Zachary JacksonAug 14, 2025
    What to do if Konica Minolta Measuring Instruments show ER11 FAILED IN CALIBRATION?
    • A
      angel47Aug 15, 2025
      If your Konica Minolta Measuring Instruments display the error 'ER11 FAILED IN CALIBRATION', it could be due to a few reasons: * Zero calibration or white calibration has not been performed correctly. Ensure that for zero calibration, the specimen measuring port is directed into the air, and for white calibration, the White Calibration Plate is used. * The Target Mask is not correct. Attach the correct Target Mask.
  • L
    lrobertsAug 19, 2025
    How to fix OK02 LOW ILLUMINATION error on Konica Minolta CM-2600d Measuring Instruments?
    • W
      William MorenoAug 19, 2025
      The 'OK02 LOW ILLUMINATION' error on your Konica Minolta Measuring Instruments can be due to: * Deterioration of xenon lamp. * Dirt on integrating sphere. Clean the integrating sphere. If the problem persists, contact a KONICA MINOLTA SENSING-authorized service facility.
  • S
    sandyrobinsonAug 27, 2025
    What to do if Konica Minolta CM-2600d displays ER05 ERROR IN FLASHING?
    • J
      jonathanreedAug 27, 2025
      If your Konica Minolta Measuring Instruments display the error 'ER05 ERROR IN FLASHING', contact a KONICA MINOLTA SENSING-authorized service facility.

Summary

Safety Information

Safety Symbols

Explains symbols used to prevent accidents during instrument use.

Manual Usage Guidelines

Provides guidelines and disclaimers regarding the manual's content.

Safety Precautions

WARNING Instructions

Details critical safety instructions to prevent death or serious injury.

CAUTION Instructions

Outlines safety instructions to prevent injury or instrument damage.

Notes on Use

Operating Environment

Specifies conditions for instrument setup and operation.

Measurement Preparation

Guidance on instrument setup, calibration, and accessories.

Power Source

Details on supplying power to the instrument.

Notes on Storage

Storage Conditions

Specifies environmental requirements for storing the instrument.

Notes on Cleaning

Cleaning Procedures

Instructions for cleaning the instrument and accessories.

Chapter 1 Before Using the Instrument

Accessories and Parts Overview

Lists accessories and identifies instrument components.

Initial Setup and Configuration

Covers preparation, system setup, and essential settings.

Conventions

Manual Conventions

Explains the structure and elements of manual pages.

Chapter 2 Preparation for Measurement

Measurement Flow and Initial Setup

Outlines measurement sequence and first-time power-on settings.

Measurement Conditions and Calibration

How to set parameters and perform zero/white calibration.

Target Data and Tolerances

Defining target colors and setting tolerances for difference measurement.

Chapter 3 Measurement

Performing and Displaying Measurements

Steps to perform measurements and interpret results.

Chapter 4 Other Functions

Averaging and Assessment

Using averaging for accuracy and assessing results.

External Connectivity and Task Mode

Connecting to devices and using guided measurement procedures.

Chapter 5 Troubleshooting

Error Messages and Solutions

Explains common error codes and their resolutions.

Chapter 6 Appendix

Technical Details and Reference

Principles, specifications, dimensions, and modes.

Konica Minolta CM-2600d Specifications

General IconGeneral
TypeSpectrophotometer
Light SourcePulsed Xenon lamp
Spectral Separation DeviceDiffraction grating
Wavelength Range360 nm to 740 nm
Spectral Resolution10 nm
Wavelength Pitch10 nm
ApertureMAV: Ø8 mm, SAV: Ø3 mm
Power SupplyAC Adapter
InterfaceUSB
Measurement Geometryd/8° (diffuse illumination, 8° viewing)
Measurement TimeApprox. 1.5 sec.
Measurement AreaMAV: ø8 mm, SAV: ø3 mm
PowerAC 100 - 240 V
Operating Temperature Range10 to 40°C
Operating Humidity Range20 - 80 %
Sphere Diameter150 mm
Inter-Instrument AgreementΔE*ab ≤ 0.2 (between two instruments)