EasyManua.ls Logo

Kroeplin C - Series - Absolute - Mode; Relative - Mode

Kroeplin C - Series
76 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32
If the „Absolute – Mode“ is activated measurements will be made with reference to the
zero point.
(This mode will be started during the initial use or after factory setting)
If the „Relative – Mode “ is activated measurements will be made with reference to the
reference value (gage block) which was selected before. This mode is used for
comparison measurement.
Switching between „ Absolute– / Relative – Mode “ can be made in MIN / MAX / HOLD
and TOL.
Switch-over from „Absolute – Mode“ into „Relative – Mode “:
The instrument must be switched on.
Finding the reference value via block gage block
(example: 10mm)
1x short
Zero point adjustment according to the reference
value
Symbol (4.1) for „Relative-Mode“ appears in the display
footer, ……
…………. as well as the reference value (example:
10mm).
If „ERR 08“ appears in the display the reference point in „Relative - Mode“ is out of the
indicating range.
Switch-over from „Relative – Mode“ into „Absolute – Mode“:
The instrument must be switched on.
Example:
Setting the zero point (10,00mm)
1x short
The reference value is activated again
Absolute – Mode / Relative – Mode

Table of Contents

Related product manuals