EasyManua.ls Logo

KYORITSU 4116A - Measurement of the Loop Impedance

KYORITSU 4116A
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
(3)VoltageMeasurement
Whenthe instrument isfirst connectedto thesystem, itwill displaythe
phase-neutralvoltagewhichisupdatedevery1s.Thismodeiscancelled
wheneverthe test button ispressed. If this voltageis not normal or as
expected,DONOTPROCEED.
5.2MeasurementoftheLoopImpedance
(1)Settheinstrumenttothe200Ω or2000Ω range.
Ifthe instrument is set to the 20Ω rangeslight sparking may occur
whentesting withthe distributionboard lead althoughthe unithas
beendesignedtominimizethis.
(2)Connecttheleadtotheinstrument.
(3)Plugthemouldedmainsplugtothesocketbeingtested.
(4)Checkthe LED's arelit as indicatedin section 5.1. Ifnot DO NOT
PROCEED-checkwiring.
(5)Notethemainsvoltageifrequired.
(6)Pressthe"PresstoTest"button.Thevalueofloopimpedancewillbe
displayedwiththeappropriateunits.Ableepwillsoundoncompletion
ofthetest.
Forbestresultsalwaystestonthelowestpossiblerange.
Forexample, a loop impedance measured on the 200Ω rangemay
giveanindicationof0.3Ω whereasonthe20Ω rangeitmayread0.28
Ω.Intheeventofthereadingbeinginexcessoftherange(e.g.more
than20Ω onthe20Ω range)theappropriateover-rangesymbol"OL"
willappearonthedisplay.
Noharmwillbedonetotheinstrumentbyselectingtoolowarange.
WARNING
IftheabovesequenceisNOTdisplayedortheREDLEDisonfor
anyreason , DO NOT PROCEED AS THEREIS INCORRECT
WIRING.Thecauseofthefaultmustbeinvestigatedandrectified.
WARNING
Thisinstrumentisintendedonlyforuseinsinglephaseoperation
at230V+10% -15%AC phasetoearth orfor useinOLD-TT
systemphasetoneutral.

Related product manuals