Mars-Mars PRO
User Manual
Code No. 9.900.900.710
U:\QA\Manualer\Arkiv\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanLaf\Mars Mars Pro manual-UK - rev. 06.doc
Page 23 af 34
Monday to Friday whole week or only a single day when the cabinet should start up in full speed
and after how long it should wait until it switches off again.
E. By pressing 8 or 9 the Time setting on the cabinet can be set by the enduser date, month, year,
clock 24 hours or AM/PM.
F. By pressing 8 or 9 Password shows up this means the user can select a Password so the keys can
not be activated without the entered Password. It can always be overruled by the Service access
code.
G. By pressing 8 or 9 Service mode can be reached. To enter this you need to be trained personnel
and to have got the Service access code.
10.1 Programming
Only for trained personnel!!
.
Service mode: This menu is only operated by authorised service people from ScanLaf A/S,
distributors and trained technical people for safety reasons who has got the Service access
code.
In Service mode can be adjusted the
Airflow speeds down flow and exhaust through 2 air flow sensors
The setting and calibration of the Alarms: analogue air flow sensors or pressostats
The configuration of the cabinet options: with or without UV light, light mode setting, el-out
lets, gas resetable valves, Pir sensor.
External control setting: output 0-10 volt to volustats, external blower dampers or CTS, input
from External blower or CTS or pressostats mounted in the exhaust channel
USB-port
Decontamination through H
2
O
2
or Formalin.
10.2 Supervision Alarm
The cabinet safety is constantly supervised so that any deviation from safe conditions or any fault in
the supervision system will be indicated immediately.
Safe conditions are indicated by a green LED light in the button for fan at normal velocity.
Unsafe conditions will be indicated both acoustically and visually by red flashing LED light.
The alarm will be initiated if
The vertical air velocity inside the work area is outside the limits.
The air volume exhausted is below the limits.
The fan is running at reduced velocity.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LaboGene SCANLAF Mars PRO-1800 and is the answer not in the manual?

LaboGene SCANLAF Mars PRO-1800 Specifications

General IconGeneral
BrandLaboGene
ModelSCANLAF Mars PRO-1800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals