The fluorescent and UV lamp (germicidal) have an average life time of about 5000 hours. SCANLAF,
however, suggests its replacement after no more than 2000-3000 hours. The luminous potential
delivered drops to 70% after said period.
REPLACEMENT OF THE FLUORESCENT LIGHT TUBES
Remove the air distributor located in top of the work chamber (handle with care)
Release the two lamps from the fixtures by rotating them 90º.
Install the new lamps performing the inverse operation.
REPLACEMENT OF THE UV lamp TUBES
The UV light is located on the back wall
Release lamps from the fixtures by rotating it 90º.
Install the new UV -lamp performing the inverse operation.
NOTE: Use gloves when manipulating a germicidal UV lamp, to prevent leaving hand prints that
reduce the effects of the ultraviolet radiation.
Replace UV lamp every 2000 or 3000 hours of life.
12.5 Spare parts
UV-light tube for Mars/Fortuna 900-1800
Light tube for Mars/Fortuna 900
Light tube for Mars/Fortuna 1200
Light tube for Mars/Fortuna 1500
Light tube for Mars/Fortuna 1800
Main HEPA filter for Mars/Fortuna 900
Main HEPA filter for Mars/Fortuna 1200
Main HEPA filter Mars/Fortuna 1500
Main HEPA filter Mars/Fortuna 1800
Main circuit board for Mars, Fortuna, Titan PCR
Light ballast for Mars/Fortuna 900 1200