Ultimate-cell Stackable Shaking Incubator V2017.2
26
Press "Stop" button in the home page to stop the current operation.
Switch the "TEMP MODE" in the "SET" page to "FIXED" mode. Press "SET" in
the home page.
Press "Fixed" in the "SET" page to start setting the general parameters (Ex-
cluding Photosynthetic Lighting)
SPEED—Speed Set button
Press it and use the keyboard to enter the speed set value, and press “OK” to confirm
and save. Set range from 30-300rpm.
TEMP—Temperature Set button
Press it and use the keyboard to enter the temperature set value, and press “OK” to
confirm and save. Set range 4.0-60.0°C.
HUMI— Humidity Set button
Press it and use the keyboard to enter the relative humidity set value, and press “OK”
to confirm and save. Set range 40-80%RH
CO2— CO2 Gas Set button
Press it and use the keyboard to enter the CO2 gas concentration set value, and
press “OK” to confirm and save. Set range 0-20%.
TIME—Timer Set button
Press it and use the keyboard to enter the timer set value, and press “OK” to confirm
and save. Set range:0-9999mins
Back—“Return” Button
Press it to exit the current page after all settings are finished.
Start the operation: Press the "RUN" button in the home page to start the fixed
value operation , the "Shaking Flask" icon will start swing. The shaking motion of
the shaking tray can be observed through the front window. The temperature,
humidity, CO2 concentration controls start functioning to reach their setpoints.
Time starts counting.
Pause the operation: During the operation, the incubator can be paused by
pressing the "STOP" button. Once it pauses, the "RUN" button keeps twinkling,
the shaking tray stops shaking motion, the timer pauses counting. But the
temperature, humidity and CO2 concentration remains under control. Press the
"RUN" button to resume the current operation under fixed value mode.
Stop the operation: When the operation is paused, press the "STOP" button
once again, it will stop the current operation. Once it's stopped, the "Shaking
Flask" stops, the "RUN" button dims, the shaking tray stops shaking motion, the
timer stops and will be reset.
The temperature, humidity and CO2 concentration controlling will also be