38
LOCK FUNC Door Lock Switch
Once the page setting is activated, press this button to switch on/off the door
locking system.
NEW PW New password Setting
Once the page setting is activated, press this button to enter the new door
locking password. The password can be set as any number within the range of
1-9999. Press "OK" in the keyboard to confirm and save the new password.
6.10.2 RS485 Communication Address Setting Option
The incubator is equipped with RS485 communication function for data logging.
To enable the function, make sure the connections between the incubator and
the communication kit is proper and secure. See individual setup manual for
detailed instructions.
Moreover, once the connection is done, communication address needs to be set
in the "LOCK/COM" page,
ADDRESSCommunication Address Setting
Press this button and enter the address of this unit for RS485 communication.
The address can be set as any number within the range of 1-250. Press "Ok" in
the keyboard to confirm and save the address.
6.11 Using USB Data Logging System
The incubator is equipped with USB interface data logging system, which
saves all the operational data automatically into mobile drive, for example,
USB drive.
Plug a USB 2.0 mobile drive into the interface on the side of the front door.
The red LED indicator shall light up during recognition or while the data is
being recorded. Do not unplug the USB drive while the red LED indicator is on.
After the first data record is saved, the system will create a folder ( NO"XXX")
in the root folder. "XXX" is based on the "Address" set in "LOCK/COM" menu.
For example, if the unit address is set as "1" the folder name will be "NO001".
All data are saved and exported as csv. format in the folder "NOxxx".
The operating data is recorded once a minute. The system can export a
maximum of 6 days of data. Once the maximum limit is reached, the earlier
data will be overwrtiten by the new data.
6.12 Using UV Light
The incubator is equipped with a UV light behind the back panel of the inner chamber.
The air is sterilized when passing though the back plenum. As it is not exposed to the
innber chamber, so it is safe to switch on the during the operation.
To switch on the UV light, press the UV light switch on the side cooling compartment.