Ultimate-cell Stackable Shaking Incubator V2017.2
27
switched off. However, if the "KEEP" in the "FUNC" page is switched on, these
functions will remain under control at their setpoints even when the operation is
stopped.
6.4.2 Operating Under Programmable Control Mode
This incubator can operate according to stored program to meet various require-
ments of professional experiments. Under programmable control mode, it can be
programmed to save up to 9 different segments, each with their own Ramp Time,
Ramp End Temperature, Hold Time and Speed. At the end of the settings, set the
Hold Deviation and Cycle Times to complete the settings.
Special Remind!
If the entered value exceeds the maximum limit, then the max value will be
set at its maximum value.
Press the “C” on the key board to clear the entry.
Setting of each parameter must be confirmed and saved by pressing “OK”
button.