EasyManua.ls Logo

Labwit ZWYC-290A - Page 50

Labwit ZWYC-290A
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ultimate-cell Stackable Shaking Incubator V2017.2
49
The platform keeps
shaking after the door
is opened
The door switch has short
circuited
Notify the distributor to replace the
door switch
As the door is
closed ,the platform
starts shaking but the
speed runs high
suddenly
Improper operating method
Refer to Users Guide and press the
Start/Stop button to operate again
Screen is not respon-
sive
Equipment is disturbed by high
frequency.
Restart the equipment-if it does not
work:
Notify the distributor.
The equipment
causes a strange loud
noise
The equipment is not placed
horizontally
Adjust the left back foot to make the
equipment stable
The fixed screw of clamp is
loose
Remove the platform and tighten the
screw
The platform is loose
Make sure the latches is securely
locking the platform in place
There is strange object, like a
piece of a bottle, under the
platform
Remove the platform ,clear the
object and clean the inside
Mechanical malfunction occurs
Notify the distributor for repair
service
The accumulation of
frost is fast after
refrigeration is started,
resulting in the rise of
temperature
The refrigerating time is too long
and the evaporating chamber is
too humid
Set the temperature to high temper-
ature, e.g. 60°C, and run for 2 hours
to dry the chamber.
Set auto-defrosting cycles.

Table of Contents