Blood Pressure:
Pressure range 0 - 300 mmHg
Accuracy +/- 4 mmHg
Calbration: Pressure sensor must be calibrated to sphygmomanometer. See Setup menu.
Pulse:
Available pulses: Carotid, Brachial and Radial
Only 2 pulses can be palpated simultaneously. (Power considerations).
Palpated BP: Radial pulse turns off at systolic BP.
Brachial pulse turns off at 20 mmHg, to prevent noise generation in
auscultation area.
Adult Child Infant
Manikin Manikin Manikin
Default Blood Pressures (mmHg) 120/80 100/70 94/66
Pulses are adjusted to BP as follows:
Systolic BP Carotid Pulse Radial/Brachial Pulse
>= 88 Normal Normal
< 88 Normal Weak
< 80 Normal Absent
< 70 Weak Absent
< 60 Absent Absent
Accessories
200-11350 Base Unit Battery Case
200-11450 Remote Control Battery Cover
200-11250 Remote Control to Base Unit Cable
200-10250 USB Cable
200-10550 DC Power Supply
200-12050 Direction for Use
200-11150 Operating Software on CD-Rom
200-10150 Carry Case