CREATE A USER TEST
1.
Press and briefly hold to turn meter on. Main Menu will appear.
2.
Press or to the Editing Menu. Press
ENTER
.
3.
Press or to scroll to the Edit User Test. Press
ENTER
.
4.
Press or to scroll to the desired test. Press
ENTER
.
Name The Test
5.
Press
ENTER
to select Name The Test.
6.
Press , and
ENTER
to select the characters in the test name.
7.
Press
ENTER
after the last character has been selected.
Select Vial/WL
8.
Press or to select Vial/WL. Press
ENTER
.
9.
Press or to scroll to the channel with the desired wavelength and vial
size combination. Press
ENTER
.
Enter 2 Point Calibration
10.
Press or to scroll to STD Calibration. Press
ENTER
.
11.
Press , and
ENTER
to select the characters in the low concentration
followed by the high concentration. Press
ENTER
.
12.
Insert the blank. Press
ENTER
. Repeat with reacted low and high concentration
standards. Press
ENTER
.
Select Decimal Places
13.
Press or to scroll to Decimal Places. Press
ENTER
.
14.
Press , and
ENTER
to select the desired number of decimal places. Press
ENTER
.
Select Units
15.
Press or to scroll to Select Units. Press
ENTER
.
16.
Press or to select the desired units. Press
ENTER
. The meter will return
to the Edit Test menu. Press
EXIT
three times to return to the Main Menu to begin
testing.
802 Washington Ave • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax: 410-778-6394 • www.lamotte.com
© 2021 LaMotte Company • 07.21 • 61910-QG-V3
®
NOTE: Use the AC wall adapter to rapidly charge the meter. Reinsert the USB port plug after charging.
WARNING: Only use the USB cable (1720-01) that is supplied with the kit. Make no substitutions.