APPLICATION: Drinking, surface, and saline waters; domestic and
industrial wastes.
RANGE: 0.0–4.0 ppm Copper
MDL: 0.1 ppm
METHOD: Cupric ions form a purple complex with bicinchoninic
acid around pH 6–7, in proportion to the concentration of
copper in the sample.
SAMPLE HANDLING
& PRESERVATION:
Copper has a tendency to be adsorbed to the surface of
the sample container. Samples should be analyzed as
soon as possible after collection. If storage is necessary,
0.5 mL of 20% hydrochloric acid per 100 mL of sample
will prevent “plating out”. However, a correction must be
made to bring the reaction into the optimum pH range.
INTERFERENCES: High concentrations of oxidizing agents, calcium, and
magnesium interfere. Silver can also interfere.
COPPER, UDV SMART3 Test Procedures 11.10
Test Procedures
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com