PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
ENTER
to select TESTING MENU.
3. Select ALL TESTS (or another sequence containing 090 Sul de LR) from
TESTING MENU.
4. Scroll to and select 090 Sul de LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use the 1.0 mL pipet (0354) to add 1.0 mL of
*Sul de Reagent A (V-4458). Cap and mix.
8. Add 6 drops of Sul de Reagent B (V-4459). Cap and mix. Wait 1 minute.
Solution will turn blue if sul des are present.
9. Use the 1.0 mL pipet (0354) to add 2.0 mL of Sul de Reagent C (4460). Cap
and mix. Color development is immediate and stable.
10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
11. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
SMART3 Test Procedures 11.10 SULFIDE, Low Range