b) Bath/Circulation thermostat
Connect 11-12 mm int. dia tubing (
Section 5.3.) to
pump connector
Pump outflow always in front, return connection always
at the back.
If the cross-section of the tubing is too small
!
temperature drop between bath
and external system due to low flow rate. Increase the bath temperature
appropriately.
Always ensure the maximum possible flow cross-section in the external circuit!
When the external consumer is placed at a higher level than the thermostat, the
pump is stopped and air penetrates into the thermostating circuit the external
liquid may drain down into the bath even with a closed system
!
danger of
flooding the thermostat!
Protect tubing with hose clips against slipping off!
When no external consumer is connected to the thermostat, the pump outflow
connection must be closed (use closing plugs) or linked to the return!