5.5 Cooling the thermostats
At bath temperatures down to just above ambient temperature (approx. 2 – 10°C) it is possible to work
without cooling. Additional cooling is required for lower temperatures
Immersion thermostat.
!
attach the cooling coil (
Section 5.1.).
Bath/Circulation thermostats
!
fitted with cooling coil, as standard..
Cooling possibilities
.
down to 20 °C Mains water
!
keep the water consumption as low as possible!
.
down to –20°C flow-through cooler DLK 10/ DLK 25 (depending on bath size and
temperature)
Section 10. Accessories
!
use water/glycol mixture (ratio 1:1).
Use insulated tubing!
When thermostating an external system the equipment must be arranged in the
following order: thermostat
!
external circuit
!
flow-through cooler
!
thermostat