EasyManua.ls Logo

Lauda Hydro H 16 A - Technical Data

Default Icon
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0
9
/2020
Hydro
Water Bath
s
25
/
34
12 Technical Data
Dimensions
H 4
340
mm
x
290
mm
x
275
mm
/ 245
mm
x
100
mm
x
165
mm
Model /
Exterior dimensions
(W
x
D
x
H
in
mm x mm x
mm)
H 8
340
mm
x
395
mm
x
275
mm
/ 245
mm
x
200
mm
x
165
mm
/ Interior dimensions
(W
x
D
x
Usable Height in
mm x mm x
mm)
H 16
500
mm
x
440
mm
x
275
mm
/ 400
mm
x
245
mm
x
165
mm
H 24
700
mm
x
440
mm
x
275
mm
/ 600
mm
x
245
mm
x
165
mm
H 41
510
mm
x
490
mm
x
476
mm
/ 410
mm
x
296
mm
x
335
mm
H 22
500
mm
x
440
mm
x
345
mm
/ 400
mm
x
245
mm
x
225
mm
H 8 A
340
mm
x
395
mm
x
345
mm
/ 245
mm
x
200
mm
x
165
mm
H 16 A
500
mm
x
440
mm
x
345
mm
/ 400
mm
x
245
mm
x
165
mm
Model / Usable Height (plus
approx. 30
mm beneath the
Water Bath
s lid)
H 4
165 mm / 114 mm /
4 l /
3,5 l
/ max water level above perforated floor up to marking "max"
H 8
165 mm / 114
mm
/
8 l /
6,9 l
/ Volume (in
litres
)
H 16
165 mm / 114 mm / 16 l / 13,7 l
/ Filling quantity from bottom of
Water Bath
up to marking “max“ (in
litres
)
H 24
165 mm / 114 mm / 24 l
/ 20,6 l
H 41
335 mm / 284
mm / 41 l / 37,6 l
H 22
225 mm / 174 mm / 22 l / 19,6 l
H 8 A
165 mm / 114 mm /
8 l /
6,9 l
H 16 A
165 mm /
114 mm / 16 l / 13,7 l
Temperature range
Display can be switched between °C and °F
Models
H 4, H
8
,
H 16, H 24, H 41
und
H 22
approx
. 5
K
above ambient to
100 °C
Models
H 8 A
und
H 16 A
approx
. 10
K
above ambient to
100 °C
Temperature
regulation
Electronic, PI
-
type
Temperature constancy
at 50
°C
+/
-
0.1
K
(temporal)
Temperature setting and display
Through four membrane keys and a 3.5" TFT
colour
display,
0.1 K
increments
Temperature cut
-
offs
Over-temperature cut-out, dependent on set point Adjustable between 0 K and +10 K,
factory setting 4 K above set point
Under- temperature cut-out, dependent on set point Adjustable between 0 K and -10 K,
factory setting 9 K below set point
Low water cut-off Electro-mechanical temperature limiter with capillary tube sensor.
Cut
-
off temperature 135 °C
-
10 K
Electrical connection
230 V +/
-
10 %, 50 / 60 Hz
Model / power
H 4 / 0,5 kW
H 41 / 1,5 kW
H 8 / 1,0 kW
H 22 / 1,5 kW
H
16 / 1,5 kW
H 8 A / 1,0 kW
H 24 / 1,5 kW
H 16 A / 1,5 kW
Mains connection
Shock
-
proof plug
, CEE 7/7
Mains fuse
,
internally
8 A, 6.3 x 32 mm fine fuse, delay
-
action
Mains fuse
, on
-
site
10 A
max.16 A
Protection
type /
protection class
IP20 / I
Ambient conditions
Laboratory rooms Use only indoors
(Not in potentially explosive surroundings)
Height
above sea level
Up to
2000 m
Ambient temperature
r
+10 °C
to
+40 °C
Humidity
Max 80
%
rel. humidity to 31
°C
, decreasing to 50
% rel. humidity at 40
°C
.
Emission
sound pressu
re level
Models
H 8 A / H 16 A
> 70 dB(A)
Model / weight
H 4 /
7,4 kg
H 41 / 21,2 kg
H 8 /
9,3 kg
H 22 / 14,9 kg
H 16 /
1
3,3 kg
H 8 A / 10,9 kg
H 24 /
17,2 kg
H 16 A / 15,2 kg

Table of Contents