EasyManua.ls Logo

Leakwise PS-220 - Page 17

Default Icon
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PS-220 RD1 USER GUIDE Revision 8 November 2017
17
The following diagram reflects the relation between sensor output signal, calibration points, set-
points, and the status shown by the front panel lights. The signal values shown are typical
values for ID-221.
Set Points & Calibration Points (µA) Status
5000µA
Water Calibration Point (R5 Gain) 1100µA (3.50V at TP3)
Water / Oil Set Point (R1, TP1) 1070µA (3.40V at TP3)
Oil / Air Set Point (R2, TP5) 950µA (~1V at TP3)
Air Calibration Point (R3 Zero) 700µA (0V at TP3)
500µA
0µA
Notes:
1. Set Point values are shown in µA (sensor raw signal) and also as volts between TP3
and TP2. All voltages are in reference to TP2 (ground).
2. Signal decrease means a thicker oil layer.
3. Oil detection sensitivity is determined by the distance between Water Calibration Point
and Water / Oil Set Point. A larger distance means detection of a thicker oil layer (less
sensitivity).
4. Oil / Air Set Point can be set as an Air indication threshold or as a Second Oil Alarm
threshold. For Air indication set TP5 to 0.5V.

Related product manuals