Leica M320 / Ref. 10 718 878 / Version 02
9.3 NOTES ON REPROCESSING OF RESTERILIZABLE PRODUCTS
Recommended: reprocess a product immediately after use.
INFO
Limitation of reprocessing
Frequent reprocessing has little eects on the product. The end of the product
life cycle is usually determined by wear and tear and damage through use.
Sterilization
Permissible sterilization methods
Article No. Designation Steam autoclave
134 °C, t > 10 min.
Ethylene oxide
max. 60 °C
10180591 Clip-on handle x
10428328 Rotary knob, binocular tube T x
10384656 Rotary knob, transparent x
10443792 Lever extension x
10446058 Protective glass, multifocal lens x
1)
10446469 Protective objective glass,
Leica M680/FL400
x
1)
10446467 Protective objective glass,
Leica M840/M841
x
1)
10445341 Handle for Leica M655, sterilizable x
10445340 Cap for Leica M655/M695, sterilizable x
10446842 Handle for Leica M400, sterilizable x
10448440 Cover, sterilizable for Leica M320 handle x
10448431 Protective objective glass, Leica M320 x
1)
10448296 Objective protective glass, Leica M720,
spare part (package of 10)
x
1)
10448280 Protective objective glass, Leica M720,
complete, sterilizable
x
1)
10448581 Cover, sterilizable for RUV800 x
1)
Products with optical components can be steam-autoclaved using the conditions listed
above. However, this may cause a layer of dots and streaks to form on the glass surface,
which may reduce the optical performance.
Cleaning
Needed: water, detergent, spirits, microber cloth
1. Flush the surface with running water (<40 °C), using a little detergent if necessary.
2. Also use spirits to clean optical components.
3. Dry optical components using a microber cloth, dry the rest of the product using
a paper towel.
Disinfection
After disinfection, thoroughly clean optical surfaces using running water/fresh drinking water
and then rinse using fresh, demineralized water.
Dry the products completely before the subsequent sterilization.
Leica Microsystems (Schweiz) AG validates:
The instructions above are suitable for preparing a product to be reused. The processor is
responsible for the desired results.
Before deviating from the instructions provided, rst verify the deviations for eectiveness
and possible consequences.
Maintenance
The Leica M320 surgical microscope is maintenance-free. To ensure operational safety and re-
liability, Leica Microsystems (Schweiz) AG recommends taking the precaution of contacting
the responsible service organization. There, periodic inspections can be agreed or a mainte-
nance contract can be concluded.