3.8 Connecting the frequency converter
Use the connector cable to attach the TURBOVAC and the TURBOTRONIK
or Turbo.Drive TD20
classic
; see the operating instructions of the frequency
converter for details.
Permissible Frequency Converters
TURBOVAC 151/151 C 600 C
361/361 C 1000 C
1100 C
Recommended Turbo.Drive Turbo.Drive
Frequency converter TD20
classic
TD20
classic
Possible TURBOTRONIK TURBOTRONIK
Frequency converter(s) NT 20 NT 20
TURBOTRONIK TURBOTRONIK
NT 151/361 NT 1000/1500
(only up to year of
production 2006
of the TURBOVAC!)
TURBOTRONIK
NT 361
Dangerous voltages are present at the mains connections. Before begin-
ning with any maintenance or service work on the pump, disconnect the
pump from all power supplies.
Operate the pump only with the matching frequency converter and con-
nector line.
Voltages of up to 400 V will be present at the connection cable between
the frequency converter and the pump; mains voltage will be present at the
fan, the flange heater, the valves and their supply leads.
Observe Safety Informations 0.2.
The connections are of the IP 40 safety classification. Do not expose the
pump, frequency converter or connectors to dripping water.
TURBOVAC 600 C,1000 C and 1100 C from year of production 2007
must not be operated with the TURBOTRONIK NT 1000/1500.