W79: Signal too small
Calibrated leak is too small or has not been opened.
Preamplifier voltage < 10 mV.
The tune process is ignored and the calibration process
is continued.
W80: Request for Cal
Request for calibration.
See explanations for menu item 24.
Since the last request for calibration, the temperature
has changed by 5 °C or the UL 200 has warmed up for
30 minutes.
W81: Factor out of Range
The calculated factor falls out of the valid range. The old
factor which has been calculated is retained.
Possible fault cause:
- The conditions for calibration have not been maintai-
ned.
- The leak rate of the calibrated leak which was entered
is much too high or much too small.
Software Problems
W84: EEPROM Write XX
The parameter could not be written into the EEPROM.
XX indicates the number of the parameter.
E85: Parameter Lost XX
At least one parameter has been lost.
The UL 200 will assume the default value for that para-
meter.
Check trigger values and calibrate the instrument. XX
indicates the number of the parameter.
W86: FB Timeout
No data have been received from the hand unit for 500
ms.
This error occurs as a follow-up error to error numbers
1 ... 5.
The hand unit has not been connected.
E87: FB CRC-Error
The data from the hand unit have not been received cor-
rectly over a period of 500 ms.
- Faulty data link.
W88: Serial Parity
See interface description SB 10.211.
W89: Serial Framing
See interface description SB 10.211.
W90: Serial Overrun
See interface description SB 10.211.
W91: Serial B. Overrun
See interface description SB 10.211.
W92: Serial Missing EOS
See interface description SB 10.211.
W93: Syntax Error
See interface description SB 10.211.
Abbreviations which are used:
FB Hand unit
MSV Mass spectrometer supply (the rear board of the
two electronics boards)
STE Control board = CPU board (the front board of the
two electronics boards)
TMP Turbomolecular pump
41
TH 10.211/8.02 - 12/97