Appendix D
D-4 Appendix D
<CELL TEMP>TRUE</CELL TEMP>
<CO2>TRUE</CO2>
<CO2ABS>FALSE</CO2ABS>
<CELLPRES>TRUE</CELLPRES>
<IVOLT>TRUE</IVOLT>
<RAW>FALSE</RAW>
</RS232>
</LI820>
Sending Data to the LI-820
To send data to the LI-820, each string must end with a '\n' to ensure that the LI-820 can parse
consecutive commands.
After data have been sent to the LI-820, the instrument replies with:
<LI820><ACK>TRUE</ACK></LI820>
if the XML was received and parsed correctly. If there was an error in the XML, the
following is sent:
<LI820><ACK>FALSE</ACK></LI820>.
For example, suppose that you would like to reconfigure the LI-820 to stop outputting cell
temperature. Here is the command to send to the LI-820: