Unpacking, Inspecting and Installing
248083408-1 Rev 1 3-3
Operation Understanding the LEDs
The front-panel LEDs have special meanings when illuminated. The
figure below shows location of the LEDs and gives a brief description
of their meaning.
Figure 3-1 Front Panel LEDs
The green POWER LED turns on when the instrument is powered
on.
The green FLOW LED turns on when the flow is correct.
The green FLOW LED will turn off if flow is insufficient.
The orange SERVICE LED will stay on steady if Laser power is
low, sensor optics are dirty or the view chamber contains foreign
objects.
The blue SAMPLING LED turns on when the instrument is
counting.
Features The REMOTE 2014P, 3014P or 5014P instrument has the following
features:
1. Optional Remote Display connection (color Touchscreen interface)
2. Four optional 4-20mA output connections
3. DIP switch control for device addressing