MODBUS Register Map v1.48
248083408-1 Rev 1 A-5
Registers 40200-40999 are reserved for character names associated
with location index values. If number of available locations is 200,
Registers 40200-40999 used for 8 character names associated with
location index. If number of available locations is 400, Registers
40200-40999 used for 4 character names.
Register 40199 indicates the number of location names supported on
this device.
Alarm Mode (40037) defines the type of calculation performed to
define an alarm condition. Alarm Mode = 0 corresponds to
conventional threshold alarming; channel bit set if threshold exceeded
for that given channel.
Alarm Parameter (40038) defines additional parameters that may be
needed in defining an alarm mode.
40203 ASCII string Location_1_char[6], char[7]
...
40996 ASCII string Location_200_char[0], char[1] (NULL terminated string)
40997 ASCII string Location_200_char[2], char[3]
40998 ASCII string Location_200_char[4], char[5]
40999 ASCII string Location_200_char[6], char[7]
400 Locations Available (Register 40050 Bit 3 = 1)
40200 ASCII string Location_1_char[0], char[1] (NULL terminated string)
40201 ASCII string Location_1_char[2], char[3]
...
40998 ASCII string Location_400_char[0], char[1]
40999 ASCII string Location_400_char[2], char[3]
Table A-3 Sensor Settings Registers
Register Data Type Description