Troubleshooting
Troubleshooting
Troubleshooting your Joy multi-user breast pump is designed for
ease of use, effectiveness, and safety. Our high standards are the
result of our certied lactation consultants’ experience with tens of
thousands of mothers expressing milk for their infants. If you have
any questions about the use or assembly of your Joy breast pump,
see “Accessory Kit Setup and Assembly” on page 19 and/or go to
limerickinc.com to make sure you have assembled the accessory kit
properly.
Problem A: 1) Pump not turning on and 2) no lights are on
are both indications of a possible power adapter failure
Instruction:
1. Ensure adapter power jack is inserted fully and doesn’t feel loose.
2. Unplug the power adapter completely from the pump and wall
outlet; inspect the cords for any signs of damage.
3. Try plugging power adapter into another wall outlet.
4. Call Limerick Customer Service.
Problem B: Not cycling; there is suction but no releases
(indication of air leak in tubing kit or possible pump failure)
Instruction:
1. Remove lter with all its tubing from pump entirely.
2. Turn on and set both vacuum and cycle speed to 30 (thirty).
3. Pump should be making a steady sound of pumping, but not a
cycling on and off sound.
4. Place nger over the small vacuum hole where the lter goes.
(Make sure you get a good seal over small opening in white