Page 34
Troubleshooting
connector for lter.)
5. You should now hear the pump sound change to cycling on and
off.
6. Disconnect the tube ends from both storage containers.
7. Put lter and tubing back into lter holder on pump.
8. Put your thumbs over both of the tube ends. (Make sure there is
a tight seal.)
9. You should now hear the pump sound change to cycling on and
off.
10. Unscrew the storage containers from the breast cup assembly
cap.
11. Inside the breast cup bottle caps are the silicone gaskets. Ensure
that the gaskets are seated well; outer ridge on gasket should be
facing you as you place into breast cup assembly cap.
12. Screw the storage container back onto the breast cup assembly
cap. Don’t over tighten.
13. Connect the tubes back onto the breast cup assembly.
14. Turn pump on, place both of the breast cups onto a hard smooth
surface, such as a tabletop.
15. Both breast cups should be compressing and releasing.
16. If pump is still not working, call Limerick Customer Service.
Problem C: Suction has become weak
Instruction:
1. Is there milk/uid in the lter? Change lter.
2. If pressure decreases and suction is poor - change lter.
3. Check for vacuum leaks. See Problem B, Instructions, steps 10-16.