Recommendations
Hot Plugging
Removing or adding any OpenBus™ cables is not allowed when the control box is on power via mains
supply or battery! If necessary anyway, follow the below procedure:
1. Remove mains or battery and wait 5 sec.
2. Mount or dismount the required cables
If this procedure is NOT followed it may result in a damaged OpenBus™ driver circuit.
The risk of a damaged circuit increases if the accessory has a high start current (in rush current).
Emergency lowering/lifting
By use of BAJ, the lifting arm can be lowered by pressing e.g. a pen in the hole or use the
control buttons, if present. This is a permitted method of lowering/lifting.
The emergency lowering/lifting “buttons” work as normal hand control buttons (you do
not get extended functionality by using these when the battery is low).
By use of BAJL, please be aware that loss of power might happen due to the battery
deep discharge protection. This will only happen by continuous use of the battery despite
warning.
Warnings
In order to avoid injury, the emergency-stop should be activated in all shipping situations.