ڦ Auto Save Function
With this function, it automatically saves the communication data as a measurement
log le in the optional SD card or USB ash drive while measuring. Since long time
communication data can be recorded, it is useful to clarify rare communication
failures of unknown cause.
Press [MENU], and then touch ‟Record control" and ‟Auto save" in order.
Save device:
Set the save destination of the log le.
Insert the SD card or USB ash drive as
selected.
Auto save
Restart
Continuous ring recording up
to the maximum number of
les.
Max-stop
Saves up to the maximum
number of files and stops
measurement.
Append Continuous ring recording
from the continuation of
the existing auto save le.
Please note that except for "Append" all the existing auto save files will be
deleted at the start of measurement.
Set the main unit memory to "Ring buer".
 
[MENU]
,
"Record control", and then "Write control" of "Capture buer".
File size
Set the size of the communication log le to be automatically saved.
Max 󳞛les
Set the maximum number of les to be recorded.
The communication log file name is # XXXXXXX.DT (XXXXXXX is a
sequential number that increases by 1 from 0000000).
When you start the measurement by [RUN], a conrmation message for overwriting
the le will be displayed. Touch [RUN] or "Continue" to start measurement. If you do
not want to lose the communication log le and want to save it to another device, press
[STOP] and save it to a PC or other device.
If the power is turned down during auto save because of low battery charged, it
may cause damage to the saved les and the storage device. Start measurement
with the suciently charged battery.