24
ڦ Auto Backup Function
Since the capture memory is DDR3-SDRAM which loses data when the power
is turned o, at the end of measurement the analyzer automatically backs up
the about 512 KB of the latest part of the measurement data into the SRAM
backed up by the battery. To automatically save the entire measurement data
or to erase the data when the power is turned o, change the setting of the auto
backup function.
Press [MENU], and then touch ‟Record control" and ‟Auto save" in order.
Auto backup:
O
No automatic backup.
Save to SRAM
Save about 512 Kbytes of the latest data into internal SRAM.
Save to File
Save all the measurement data to the storage device
The data saved in the SRAM is automatically loaded into the capture memory when the
power is turned on.
The le automatically saved in the storage device, @AUTOBUn.DAT (n is the number of the
used capture memory - BUF0, BUF1, or BUF2), needs to be loaded manually to use.
ڦ Auto Run/Stop Function
By specifying the date and time for the start and end of measurement, it
automatically executes measurement for the specied period.
Press [MENU], and then touch ‟Record control" and ‟Auto run" .
Mode:
Select the repeat condition for automatic measurement.
Touch to activate the checked item.
Set the automatic measurement
period for ‟Run time" and ‟Stop
time".
With the settings on this screen,
measurement starts at 12:30 every
day and ends automatically at 13:00.
If you want to start the measurement in conjunction with the power on of the target line,
check "External power on Run".