ڦ Communication Condition Auto Setting Function
The monitor processing of the analyzer estimates the communication
conditions of unknown communication lines for measurement.
By touching “Auto Conguration in the [ ] menu, Analyzing transmission
speed...is displayed and the analysis process of communication condition
starts.
When the communication conditions are determined, the details will be
displayed. A few seconds after that, the current communication condition
settings are automatically changed to the determined communication
conditions and it starts the measurement.
If you touch Cancel” or press [STOP] or [ESC] before it starts measurement, the analysis
processing of communication conditions will be stopped and the current communication
condition settings will not be changed.
The following conditions are required for the line to be measured in order to judge the
communication conditions correctly.
Communication speed is 460.8Kbps or less.
Protocols are ASYNC, SYNC/BSC, or HDLC/SDLC.
On the line communication data without error is owing at a certain frequency.
On the line communication data including bit patterns of ‘101’ or ‘010’ is owing.