(7) ...... Press key to advance to ‘Response Rate’. This is
programmed inpH units.
The smoothing of the input signal is determined by delta
(pH) time. The ‘Response Rate’ is the time that the
computer display takes before it updates the pH readings.
When 00 : 01 is programmed the controller responds to a change in input in one (1) second.
If 00 : 10 is selected the controller responds to a change in input in ten (10) seconds. (i.e., the value displayed is the average
of the 10 previous 1 second readings.)
Examples: The sampling time (delta) is 00 : 10 and the current reading is 2.00 pH.
When the pH input is increased instantaneously to 12.00 pH, the display will respond as follows:
Press the key to advance to set temperature.
This setting is relevant when no temperature probe
(1000 platinum RTD) is connected.
If the sampling was 00 : 01 seconds, the response would be:
Display Key:
While in the Run Mode the
key can be pressed once to display current parameters. Each screen will come up for
three (3) seconds and then returns to pH display or System Run automatically (screens shown below are: mV, °C, and mA).
Seconds 0 1 234567891011
pH 2345678910111212
Seconds 0 1 234567891011
pH 21212121212121212121212
The following values can be programmed (min : sec):
00 : 01 00 : 10 00 : 20 00 : 30. . . . . . 04 : 00 (In increments of 10 seconds)