7 - Faults
Page 14
7.0Faults
Error Message: "Prog"
The instrument memory has been corrupted. You can attempt a "Power on
reset".
1. Switch off the power supply.
2. Press and hold all three center buttons and switch the power on. If the instrument
can recover itself, it will perform a short self test routine and return to the normal
display. If not contact your local Loup Electronics dealer.
NOTE: All calibration settings, sub-totals, grand total and weighing units will be reset
to the factory default settings.
Error Message: "Err"
If the display shows "Err" there is no pressure signal being received by the scale.
This indicates either a bad connection, wiring or a faulty pressure sensor.
Visualy inspect the wiring and connections first, then if necessary check
continunity of the sensor lead wires between the Junction Box terminals and the
3 pin weatherpack connector on the load sensor extension cable. If no fault is
found contact your local Loup/RDS dealer.