7 - Faults
Page 14
7.0Faults
Error Message: "Prog"
The instrument memory has been corrupted. You can attempt a "Power on
reset".
1. Switch off the power supply.
2. Press and hold all three center buttons and switch the power on. If the instrument
can recover itself, it will perform a short self test routine and return to the normal
display. If not contact your local Loup Electronics dealer.
NOTE: All calibration settings, sub-totals, grand total and weighing units will be reset
to the factory default settings.
Error Message: "Err"
If the display shows "Err" there is no pressure signal being received by the scale.
This indicates either a bad connection, wiring or a faulty pressure sensor.
Visualy inspect the wiring and connections first, then if necessary check
continunity of the sensor lead wires between the Junction Box terminals and the
3 pin weatherpack connector on the load sensor extension cable. If no fault is
found contact your local Loup/RDS dealer.