Page #33 background imageLoading...
Page #33 background image
33
CheckitDirect_Cl-Br-pH-CyA-TA-CaH_2 09/2010
pH-value with liquid reagent
6.5 – 8.4
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see “Operation").
Fill the vial with drops of the same size by holding the bottle vertically
and squeeze slowly:
6 drops of PHENOL RED solution.
Close the vial tightly with the cap and invert several times to mix the
contents.
Place the vial in the sample chamber making sure that the
marks
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display as pH-value.
Tolerance: ± 0,2 pH
0.0.0
PH
Zero
Test
GB
Methods
RESULT
Notes:
When testing chlorinated water the residual chlorine contents can influence the colour 1.
reaction of the liquid reagent. This can be avoided (without interfering with the pH
measurement) by adding a small crystal of Sodiumthiosulfate (Na
2
S
2
O
3
· 5 H
2
O) to the
sample before adding the PHENOL RED solution. PHENOL RED tablets already contain
Thiosulfate.
Due to differing drop sizes results can show a discrepancy in accuracy by comparison 2.
with tablets. This can be minimised by using a pipette (0.18 ml PHENOL RED solution is
equivalent to 6 drops).
After use replace the bottle cap securely.3.
Store the reagent in a cool, dry place ideally between 6°C and 10°C.4.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lovibond CheckitDirect and is the answer not in the manual?

Lovibond CheckitDirect Specifications

General IconGeneral
BrandLovibond
ModelCheckitDirect
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals